Why Is Uniform Illumination Harder Across a Full-Size PV Module?
When a full-size PV module is placed inside a solar simulator, how evenly are its centre, corners and edges illuminated?
For laboratories selecting or accepting equipment, the answer requires measurements across the area they will actually use. Module dimensions, positioning and changes during extended operation all matter.
Drawing on factory commissioning records from Zealwe Tech’s Steady-State Solar Simulator, this article examines lamp arrangement, edge coverage and operating conditions, with practical points for equipment selection, acceptance and maintenance.

PV module positioned inside a Zealwe Tech Steady-State Solar Simulator.
As the illuminated area expands from a small specimen to a full-size module, edges, boundaries between lamp arrays and obstructions become more significant.
Adding lamps or increasing power changes the illumination. To determine whether it becomes more uniform, engineers need to examine the irradiance distribution across the entire test plane.

Position of the overhead lamp assembly relative to the illuminated test area.
1. Understanding the Uniformity Figure
Start with the test the equipment needs to perform.
UV ageing involves irradiance within a specified wavelength range, accumulated dose, temperature and humidity. Continuous illumination also requires attention to specimen temperature and its condition after prolonged exposure.
“Steady-state” describes continuous illumination. A solar simulator can operate in this mode.
IEC 60904-9 evaluates solar simulators by spectral match, spatial non-uniformity and temporal instability. An AAA classification means Class A in all three categories. The 2020 edition also introduced Class A+ and extended-wavelength spectral evaluation.[1]
Spatial non-uniformity is calculated using the following formula.
N = (Emax − Emin) / (Emax + Emin) × 100%
Emax and Emin are the highest and lowest irradiance values within the specified test area. A lower N means a smaller relative difference between these extremes.

Spatial non-uniformity classes and illustrative irradiance values.
The examples use (Emax + Emin) / 2 = 1000 W/m². For example, minimum and maximum readings of 980 and 1020 W/m² give N = 2%.
These values illustrate the calculation. The measurement area and operating conditions should accompany any actual test result.
When reviewing equipment specifications, laboratories should check the area, wavelength range and operating conditions associated with the uniformity figure, then compare them with their own testing requirements.
2. How Individual Lamps Build the Overall Distribution
Light from a single lamp reaches different points on a plane at different distances and angles. Without additional homogenising optics, the centre is often more strongly illuminated than the edges.
In a multi-lamp system, each point receives contributions from several lamps and reflected light. Lamp spacing, distance to the test plane, emission direction and reflector geometry determine how these contributions overlap.
Insufficient overlap can leave low-irradiance areas. Excessive concentration can create local peaks.

Overlapping contributions from individual lamps form the irradiance distribution across the test plane.
If increasing power raises every reading by the same proportion, N stays the same. Evaluating an adjustment therefore requires comparing the relative readings across the entire area.
3. A Larger Test Area Brings the Edges into Focus
Larger specimens, or a wider placement area, may extend into regions with weaker illumination. With limited installation space and available power, extending coverage can make these differences more apparent.
Three locations deserve particular attention.
Outer edges, where fewer neighbouring lamps contribute light.
Boundaries between lamp arrays, where overlapping illumination must be balanced.
Areas near supports or specimens, where structures may block part of the light path.
Chamber-wall reflections can also change the local distribution. Additional edge illumination must work with the existing light field.

Edge coverage, lamp-array boundaries and structural shading.
The measurement grid should cover the intended specimen position, including corners and areas near potential obstructions. If the placement changes, these locations should be checked again.
A map showing measurement positions and irradiance readings helps reveal where the highest and lowest values occur, and how they relate to the module footprint.

Measured irradiance distribution across the test plane.
4. Operation and Maintenance Can Change the Distribution
Lamps, reflectors and optical materials warm up at different rates. Temperature affects lamp output, while thermal expansion can alter component positions or angles.
Over longer periods, lamp ageing, dust and changes in optical materials also affect performance. If these changes differ between lamp positions, the original distribution can shift.
During factory commissioning of a Zealwe Steady-State Solar Simulator, engineers exchanged the filters at lamp positions 2 and 3 and compared the readings.
Both records, dated 16 and 21 October, listed a test temperature of 60°C, a power setting of 90% and the same notation for 110 hours of operation at 85°C/85% RH. The later record documented the filter exchange.[2]

UVB percentage at two lamp positions before and after the filter exchange.
UVB percentage was calculated from the portable radiometer readings using UVB / (UVA + UVB) × 100%.
At position 3, the UVB percentage fell from 5.49% to 2.53%, a decrease of 2.96 percentage points. At position 2, it rose from 2.41% to 4.06%, an increase of 1.65 percentage points.
With the same power setting, the two positions changed in opposite directions. Together with the filter-exchange record, this gave engineers a reason to examine the relationship between each filter and its lamp position.
This individual-lamp comparison helps investigate differences in wavelength composition. Evaluating spatial uniformity across the full illuminated area requires repeat measurements in the target wavelength range using the original measurement grid.
5. Each Adjustment Has a Trade-Off
Improving uniformity involves extending coverage, adjusting overlap or redistributing light. The practical choice depends on target irradiance, installation space and operating and maintenance requirements.

Approaches to improving illumination uniformity and their practical trade-offs.
These trade-offs also affect equipment selection. Specifying module dimensions, placement, target irradiance and available space alongside the uniformity requirement helps establish whether a proposed configuration fits the laboratory’s needs.
6. What to Check in a Measurement Report
Measurement conditions
Check the wavelength range, irradiance, effective illuminated area, power setting, temperature and operating duration.
When comparing equipment, check that the reported results were obtained under corresponding conditions.
Measurement positions
Compare the measurement grid with the intended module position. Review the individual readings and the locations of the highest and lowest values.
A uniformity percentage and a distribution map together help identify weaker areas and whether they fall within the specimen footprint.
Measurements after warm-up and maintenance
Review records taken after thermal stabilisation.
Following lamp replacement, cleaning or changes to optical components, repeat measurements at the original positions and operating conditions. This helps assess whether the distribution has been restored and provides a basis for tracking subsequent changes.
From Factory Measurements to Laboratory Use
Designing and commissioning a steady-state solar simulator requires coordinated work on light sources, optical components, mechanical arrangement and operating conditions. Measurements show how these elements perform together.
The factory records shared here illustrate part of that work at Zealwe. For laboratories, comparing measurement records with their own module dimensions and test conditions provides a practical basis for equipment selection, acceptance and subsequent maintenance.
To discuss a steady-state solar simulator for your laboratory, share your module dimensions, target irradiance and test conditions with the Zealwe team.
Explore Zealwe’s PV testing equipment